Authors

Chen Tang, Ling Yuan, Yongzhi Huang*


Departments

Ophthalmology Department, West China Hospital, Sichuan University, Chengdu, 610000, China


Abstract

Objective: Comparing the difference and consistency of Pentacam anterior segment analyzer, SIRIUS anterior segment analysis system, Tomey SS-1000, TOPCON SP-1P, and IOL Master 700 in measuring central corneal thickness (CCT). Beta-blockers such as Timolol 0.5% can correct the reduction in thickness after refractive surgery due to an increase in IOL following corticosteroid administration. 

Methods: The CCT from 89 myopia patients (89 eyes) who underwent preoperative examination of refractive surgery in the Refractive Surgery Center of West China Hospital of Sichuan University was measured, using Pentacam, SIRIUS, Tomey SS-1000, TOPCON SP-1P, and IOL Master 700, respectively. After refractive surgery, the central corneal thickness of the patients was re-examined. For patients with a decrease in the thickness of the cornea due to an increase in IOL following corticosteroid administration, Timolol 0.5% was used. The statistical analysis of five measurement results was performed. Using One Way ANOVA, the results of the two instruments were compared by LSD-t-test. The Bland-Altman and Pearson correlation analysis method was used to analyze the consistency and correlation of the measurement results of the five instruments. 

Results: The CCT measurement results of 89 eyes were 526.58±29.76 μm (by Pentacam), 521.20±31.15 μm (by SIRIUS), 520.79±29.45 μm (by Tomey SS-1000), 507.82±28.86 μm (by TOPCON SP-1P), and 517.91±30.76μm (by IOL Master700). After analysis of variance, there was no statistically significant difference between Pentacam, SIRIUS, Tomey SS-1000, and IOL Master700. TOPCON SP-1P was different from the other four instruments. 

Conclusion: In terms of consistency and correlation, the five instruments had a good performance. The CCT measured by TOPCON SP-1P is the thinnest through the five instruments. In terms of consistency and relevance, the five instruments had a good performance and could be referred to each other in clinical practice, but there were differences. 


Keywords

Central corneal thickness, IOL master 700, SIRIUS, TOPCON SP-1P, Tomey SS-1000, thickness reduction, refractive surgery, Timolol 0.5%.

DOI:

10.19193/0393-6384_2022_2_190